How to Fill Out the Amico Floor Pedestal Device and Accessory Chart: A Step-by-Step Guide for Healthcare Facilities
Configuring a medical gas floor pedestal for an operating room, surgical suite, or ambulatory care center is a detailed process. The Amico floor pedestal is a popular choice, but its specification sheet, specifically the "Device and Accessory Chart" on page two, can be confusing. A mistake here can lead to ordering the wrong equipment, causing project delays and costly re-orders. This guide provides a definitive, step-by-step walkthrough of each field on the chart to ensure you get your configuration right the first time. We'll cover everything from gas outlet styles to specialized electrical devices, helping you specify a pedestal that meets both your clinical needs and strict regulatory standards.
Understanding the Basics: Pedestal vs. Boom
Before diving into the chart, it's important to understand the role of a floor pedestal. Unlike an equipment boom or ceiling column, which is mounted overhead, a floor pedestal is a fixed-base unit that brings essential services, medical gases, electrical power, and data, from the facility's infrastructure to the patient care vicinity. They are common in areas where ceiling heights are low or the structure cannot support a boom. The Amico 12" and 18" Floor Pedestal is a robust solution for these applications.
Gas Outlets by Style: The Core of Your Pedestal
The first and most critical section of the chart is for specifying the medical gas outlets. These are the connection points where clinicians hook up flowmeters, ventilators, and other essential equipment. The chart asks you to fill in the quantity for each gas type based on the connection style.
Outlet Styles: DISS vs. Chemetron
Amico offers several connection styles, but the most common in the United States are DISS and Chemetron.
- DISS (Diameter Index Safety System): This is the most prevalent standard. DISS connectors use a series of non-interchangeable threaded connections, ensuring that a medical air hose cannot be accidentally connected to an oxygen outlet, for example. This is a core safety principle outlined in NFPA 99.
- Chemetron: Also known as NCGA (National Compressed Gas Association), this style uses a latch-key mechanism with distinct shapes for each gas service. They are also non-interchangeable.
Your choice will depend on the existing standard used in your facility. For new construction, DISS is generally the default.
Filling out the Gas Outlet Section
To fill out this part of the form, you will enter the number of outlets you need for each gas type (Oxygen, Medical Air, Vacuum, etc.) in the column corresponding to your chosen style (e.g., "DISS").
- Example: If you need two oxygen outlets and one vacuum outlet, you would write "2" in the Oxygen-DISS box and "1" in the Vacuum-DISS box.
Why this matters for compliance: NFPA 99, section 5.1.3.1, mandates the use of gas-specific connectors to prevent interchangeability between different pipeline systems. Using a consistent, indexed system like DISS or Chemetron is not just a choice, it's a fundamental requirement for patient safety. After installation, a certified ASSE 6030 verifier from a company like Medical Testing Solutions must confirm that each outlet provides the specified gas and that the indexing system is functioning correctly as part of the medical gas verification process.
Electrical Devices: Powering Your Equipment
The next section covers electrical receptacles. It's divided by color (Red vs. Ivory/White) and type.
Red vs. Ivory Receptacles
- Red Receptacles: These signify that the outlet is connected to the hospital's emergency power system or essential electrical system, as defined by NFPA 99, Chapter 6. They will have backup power from a generator in the event of a utility outage.
- Ivory/White Receptacles: These are connected to the normal utility power.
All critical patient care equipment should be plugged into red outlets.
Types of Electrical Devices
You'll need to specify the quantity of each type of receptacle you need.
- 15A or 20A Duplex: This is your standard NEMA 5-15R or 5-20R outlet.
- GFCI (Ground Fault Circuit Interrupter): GFCI outlets are required in "wet procedure locations" as defined by the NEC (National Electrical Code). They provide an extra layer of protection against electric shock. If the pedestal is in an OR or another area where liquids are present, you'll need GFCIs.
- Twist-Lock Receptacle: These are NEMA L-series connectors that lock the plug in place, preventing accidental disconnection of critical equipment.
- Ground Jack: Provides a central, reliable grounding point for various pieces of equipment to prevent static buildup and ensure electrical safety.
- Monitor Jack: A specialized connection point, often for patient monitoring systems.
When filling out the chart, enter the quantity of each device you need under the appropriate color (Red or Ivory).
Data and Communication Provisions
Modern healthcare relies on data. This section allows you to specify empty provisions for data jacks. You'll typically have your IT or communications contractor pull the cable and terminate the jacks. The form asks for the number of "Data Provisions (Specify)," where you might write "2-Cat6" or "1-Coax," depending on your needs.